• Westergren ESR Test Kit
  • Westergren ESR Test Kit

Westergren ESR Test Kit

Westergren ESR Test Kit

KANGJIAN

Product Overview
  • Westergren ESR Test Kit

  • Overview

  • KJ823‑2 Westergren ESR Rack, Designed for use with Type B Westergren ESR components (KJ823‑1 and KJ823‑3).

  • Westergren ESR Measuring Tubes, Manufactured from highly transparent polystyrene (PS) material, featuring clear graduations for precise reading by medical personnel.


  • How to Use

  • ① Sample Preparation

  • Type A: After blood collection using Ø13×75 mm ESR vacuum tubes, gently invert the tube 8‑10 times to mix (avoid bubbles)
  • Type B: For syringe‑drawn samples, transfer blood to KJ823‑3 container and invert 8‑10 times.Place the tube vertically in the ESR rack.

  • ② Insert Measuring Tube

  • Slowly insert the Westergren ESR measuring tube into the sample tube. *Type A: Remove the vacuum tube cap prior to insertion. Ensure the tube tip is immersed in the blood sample.

  • ③ Adjust to Zero Mark

  • Gently press the measuring tube downward until the blood level aligns with the "0" graduation mark.

  • ④ Reading Result

  • Let stand undisturbed at room temperature for exactly 1 hour.Report the erythrocyte sedimentation rate in mm/hr (millimeters of plasma column height)
Specifications
REF.

Outer Dimension

MAT.
Qty/BagQty/BoxQty/CsDetailed information
Original REF.
KJ823-2

Rack:185×45×42mm

PS1-30Type:B179-0015
KJ823-1
ESR Tube:Ø7×220mmPS502004000179-2400
KJ823-3Blood Tube:Ø14×58mmLDPE1004004000179-2410


Label:
Product Video
Label: