• Westergren ESR Test Kit
  • Westergren ESR Test Kit

Westergren ESR Test Kit

Westergren ESR Test Kit

KANGJIAN

Product Overview
  • Westergren ESR Test Kit

  • Overview

  • KJ822 Westergren ESR Rack, Compatible with Type A Westergren ESR Measuring Tubes (KJ823) and Ø13×75 mm ESR vacuum blood collection tubes.

  • Westergren ESR Measuring Tubes, Manufactured from highly transparent polystyrene (PS) material, featuring clear graduations for precise reading by medical personnel.


  • How to Use

  • ① Sample Preparation

  • Type A: After blood collection using Ø13×75 mm ESR vacuum tubes, gently invert the tube 8‑10 times to mix (avoid bubbles)
  • Type B: For syringe‑drawn samples, transfer blood to KJ823‑3 container and invert 8‑10 times.Place the tube vertically in the ESR rack.

  • ② Insert Measuring Tube

  • Slowly insert the Westergren ESR measuring tube into the sample tube. *Type A: Remove the vacuum tube cap prior to insertion. Ensure the tube tip is immersed in the blood sample.

  • ③ Adjust to Zero Mark

  • Gently press the measuring tube downward until the blood level aligns with the "0" graduation mark.

  • ④ Reading Result

  • Let stand undisturbed at room temperature for exactly 1 hour.Report the erythrocyte sedimentation rate in mm/hr (millimeters of plasma column height)
Specifications

REF.Outer DimensionMAT.Qty/BagQty/BoxQty/CsDetailed informationOriginal REF.
KJ822Rack:183×57×150mmPS1-30Type A179-2300
KJ823ESR Tube:Ø4.5×230mmPS502004000179-2300


Label:
Product Video
Label: